HSMP Equipment & Furniture Component HPIU: HH
equipment specifications[1].doc 16/03/2011
page 1 of 27
STANDARD LIST OF MEDICAL EQUIPMENT & THEIR TS
Item
No
Name Quantity Technical
Specifications and Standards
1
X-Ray Film processor
tabletop
1
Processing machine for X-ray films from 13x18 cm. 18x24 cm.24x30
cm.30x40cm.35.6x35.6 cm. 35x43cm.
Capacity at least 60 films per hour
Standard Equipment: Processor + set of 3 replenish bottles
Accessories: Water supply connection with shut-off valve , Filter , Light-tight
cover , Processor stand
Automatic stand-by mode.
Automatic film detection
Power Requirements :
220 VAC ±10 % 50 Hz .
2 X-ray Film Viewer
8
X-ray Film Illuminator (Viewer) .
2 Fields . Wall-Mount
Diffuse, uniform, flicker free illumination
Transparent Spring-loaded film Retainers shall grip lightly and firmly without
obscuring top edge details
Screen shall be recessed into the cabinet to help keep the interior dust free and
eliminate side light spill
The film retainers shall always operate effectively maintaining an even pressure
across the full width of the illuminator.
Direct Starting , Luminous Source : Daylight
Command and Control on Front Side.
Bipolar Switch with Pilot lamp
Wall Fixations shall be provided for . Cord , Local Plug
Power Supply : 220V/50Hz
CE;
EC Marked
US FDA;
ISO certification
3 X-Ray Unit Universal
1
Universal, Remote controlled Universal Unit
Processor controlled, Mixed Cassette’s Radiography and Digital Fluoroscopy
screening System.
Overhead Tube on the freely moving Tube Arm without floor mounted column.
The System to be suitable for Standard Skeletal and Radiographic Examinations,
Including Lateral Exposures and Oblique Beam Projections
Fully Automatic Under Table Spot Film Device for Cassette Radiography with
Extensive Range of Cassettes, Grids and Attachments for Image Intensifier.
Free Cassette Exposures on Table, Floor, Wheelchair or Gurney.
Table.
Tilt: Motor Driven, +90º to -15º, automatic stop in horizontal position
Height: approximately 85 cm.
Tabletop Outside Dimensions: length approximately 200cm,Width approximately
80cm
Radiolucent: not less than 190cm x 55cm.
Longitudinal Travel: Motor Driven, at least 160 cm
Transverse Travel: at least 20 cm
Patient Weight: at least 200 kg.
Tube Assembly.
Max. exposure voltage : 150 kV
Anode heat dissipation rate at least 30 kW for small focus and 60 kW for large
focus
Anode heat storage capacity at least 500 kHU
Dual focus 0.6 mm and 1 mm
Complete filtration W _ 2.5 mm Al
Focus-Film Distance (SID) Fixed Variable : 115cm / 150cm. 115cm and 150cm
must be set by Motor Driven with Adjustment Speed
Oblique Projections: max.-40º-+40º ( SID 115cm.) and -35º-+35º (SID 150cm),
Automatic Parallax Compensation Between Cassette and Image Intensifier Input
Screen in Central Ray.
HSMP Equipment & Furniture Component HPIU: HH
equipment specifications[1].doc 16/03/2011
page 2 of 27
Tube Assembly Swivel: Manually in the Range +90º to -90º with Stops max. every
10º and -90º to 180º with stops max. every 30º.
Spot Film Device.
Front loading, automatically drawing in and out, centering and format sensing for
Cassettes, of the Formats 18x24cm (8"x10") to 35 x 43cm (14"x17"), Standardized
According to IEC, ANSI and DIN
Automatic Format Collimation: must be Selected Separately According to Format
Height/Width for Cassette Spot filming; Automatic Formatting for Bucky
Exposures, Object-related Collimation.
Film Segmentation: Minimum 4 on 1., in Multiple Segmentation According to
Cassettes Program
Inward Movement Time: Park to Exposure position not more than 1sec.
Time Interval: Fluoroscopy/Radiography _1.0 sec. without grid movement.
Scattered Radiation Grid: Stationary 17:1, 70 Lines/cm, Fo=125cm.,
Excursion: minimum 105cm, Remote-controlled.
Compression Device.
Radio transparent, Remotely Controlled, Detachable and Replaceable Cone
minimum 3 Shapes.
Compression Force range: from 5 to 155N, Movement Blockage starting from 50N
Force Indication: Digital at the System Remote Control Console.
Projection Angle: -30º-+30º.
Image Intensifier and TV System
3 Image Fields 13 to 23 cm
Visual resolution: Mean Value 5.2; 5.8; 6.4 Lp/mm
Television System Standard: Line Frequency 50Hz; 625 Lines with 50Hz.
TV Technology : CCD sensor, noise suppression, LIH function
Automatic Dose Stabilization (ADR)
Monitor: 44cm or more, Frame Rate at least 50 Hz.
X-Ray Generator
High Frequency, Multipulse
Up to 150 kV, 50kW.
Automatic System: 1-Point Technique with Continuously Falling Load; 2 -,3- Point
Technique with Constant Load; 3-Point Technique with AEC
Organ Programs: minimum 20 Organ Programs must be generated and stored.
Number of Workstations: not less than 5 of with 1 Fluoroscopic Workstation
Fluoroscopy working range: minimum 40kV/ 0.5mA up to maximum
110kV_10mA.
Fluoroscopic Mode Selectable: Fluoroscopy Under Manual Control / Automatic
Control
Fluoroscopy Under Manual Control: manual select kV but mA must be calculated
from Anti-isowatt, curve
Fluoroscopy Under Automatic Control: Minimum 2 required Characteristic Curve
must be selected for Automatic Dose Rate Control : us Minimal Dose/ Contrast
Litho ( or other according customer ordering)
Tube Connection: minimum 2 Double-Focus Tube Assembly.
3 field Automatic Exposure Control
Primary Collimator: Inherent Filtration 1.0mm Al; additional Filters 0.1mm Cu,
0.2mm Cu, 0.3mm Cu.
Ambient Conditions
Operation: +10Cº to 35Cº; 20% to 75% relative humidity, non-condensing; 70kPa
to 106kPa
Storage and Transport: -20Cº to+70Cº; 10% to 95% relative humidity, noncondensing;
70kPa to 106kPa
Power Connection
Nominal Voltage: 3/N/PE, 380V±10%
Nominal Frequency: 50Hz.
Completeness:
Lead glass, radiation protection window 80x100cm. ±10 % , 2,1 Pb equiv
X-ray protection:
1x Coat apron, 0.35 Pb, size medium, 110cm. X-ray protection coat apron for
general examinations.
1x Gonad protection apron 40x37cm, 0.5 Pb. Apron for adults for
protection of ovarian and gonad.
1x Gonad protection capsule for boys, 1.0 Pb.
- Set of cassettes
- Cassette without patient ID window
- Fitted with green screen
2x Cassette with green screen, 18 cm x 24 cm
HSMP Equipment & Furniture Component HPIU: HH
equipment specifications[1].doc 16/03/2011 page 3 of 27
1x Cassette with green screen, 18 cam 43 cm
2x Cassette with green screen, 24 cm x 30 cm
2x Cassette with green screen, 35 cm x 43 cm
1x Cassette with green screen, 35 cm x 35 cm
4 X-Ray Mobile
1
Exposure range at least 0.5–200 mAs
Shortest exposure time: not more than 10ms.
Max. exposure time: not less than 5 sec.
Tube head must be rotatable in all planes.
Adjustable collimator and light marker Easy to use and to move. Non automotive
Two point technique – kV and mAs.
High- frequency X-ray generator with multipulse voltage provider.
The touch-screen keys and digital display of kV and mAs values.
Output: not less than 2.5kW. 120 kV, 40mA.
Focal spot not more than 1 mm.
Collimator: rotatable ± 90°
Max. horizontal extension at least 800mm
Power Requirements
220 V ±10%, 50Hz
5 Ultrasound (3 probes)
1
Wide Dynamic Range Digital Front-end
The 12-bit A/D converter
Multi-beam processing.
Image Display Modes
B: gray-scale imaging, M,D: Spectral Doppler (PW,HPRF PW),
Flow: Color Doppler and Power Flow imaging, Dual B, B and M, B and D, B(Flow),
Dual B (Flow), M(Flow), B(Flow) and M(Flow), B(Flow) and D,
Triplex mode: B, Flow and PW Doppler simultaneous real-time display,
Dual Dynamic Display (DDD): B and B (Flow) simultaneous real-time display, TDI
(Tissue Doppler Imaging)-Available on electronic scanning probes.
Request function: In multi-mode display, it is possible to select one image for full
screen display.
DMS (Integrated Data Management System)
Measurement and Analysis.
General measurements
On B-mode image
On M-mode image
On spectral Doppler
On B/D-mode image
On B(Fiow) mode image
Obstétrical measurements & calculations
Gynecological measurements & calculations
Peripheral vessels analysis
Urological measurements & calculations
Abdominal measurements
Report Functions
Data Communication Function
Preset Function
Characters and graphic displays
Probe connectors
Standard
Active connectors: 2 connectors
Dummy connector for rest: 1 connector
Viewing Monitor
15-inch diagonal color Liquid Crystal Display with brightness and contrast controls
Acoustic Power
0 to 100%, changeable
Environmental Requirements
Temperature: +10 to +40 degrees C
Relative Humidity: 30 to 75%
Power Requirements: 220V / 50 Hz.
Probes
General abdomen OB/GYN 2.5-6.0 MHz.
Small parts, PV (steered linear) 5.0-10.0 MHz.
Adult heart (harmonic echo) 2.1-3.8 MHz.
6
Gastroscope (With halogen
light source)
Optical System :
Field of view: at least 100
Depth of field: at least 4 to 100 mm
HSMP Equipment & Furniture Component HPIU: HH
equipment specifications[1].doc 16/03/2011
page 4 of 27
1
Distal end : Outer diameter: not more than 10mm.
Bending Section
Range of tip bending: not less than Up 210, Down 90, Right 100, Left 100
Insertion tube : Outer diameter: not more than 13.3mm
Working length: not less than 1025mm.
Length markings
Instrument channel :
Inner diameter: not less than 2.8mm.
Light source
Halogen technology shall provide a high light intensity of at least 150W.
Variable light intensity.
Fast and easy bulb change during operation : In case of a bulb failure immediate
switch to the second bulb with the reverser shall be possible.Adjustment of
brightness and power with buttons and switches on the front panel.
Shall be connectable with conventional telescopes, fiberscope.
Shall be adaptable to all endoscopic procedures
Optionally , an automatic built-in swivel mechanism will bring the spare halogen
bulb into working position.
Technical Data : Halogen , 150 watts , Halogen, 2 pieces
Manual switch , 2 x 150 W , 24 V
Power Supply : 220 VAC 50 Hz .
Standards
CE;
EC Marked
US FDA;
ISO certification
7
Colonoscope
(With halogen light
source)
1
Optical System :
Field of view: at least 120
Depth of field: 5 –100 mm.
Distal end : Outer diameter: not more than 13mm.
Bending Section
Range of tip bending: not less than Up 180, Down 180, Right 160, Left 160
Insertion tube : Outer diameter: not more than 13.3mm
Working length: not less than 1680mm.
Instrument channel
Inner diameter: not less than 3.2mm.
Light Source
Halogen technology shall provide a high light intensity of at least 150W .
Variable light intensity
Fast and easy bulb change during operation : In case of a bulb failure immediate
switch to the second bulb with the reverser shall be possible
Adjustment of brightness and power with buttons and switches on the front panel.
Shall be connectable with conventional telescopes, fiberscope,
Shall be adaptable to all endoscopic procedures
Optionally , an automatic built-in swivel mechanism will bring the spare halogen
bulb into working position.
Technical Data : Halogen , 150 watts , Halogen, 2 pieces
Manual switch , 2 x 150 W ,24 V
Power Supply : 220 VAC 50Hz .
Standards
CE;
EC Marked
US FDA;
ISO certification
HSMP Equipment & Furniture Component HPIU: HH
equipment specifications[1].doc
16/03/2011 page 5 of 27
8
Electrocardiograph 3
channels
3
Recording ECG Leads: 12 standard Leads
Recording Channels: 3/1 user selectable
LCD display of ECG
Lead switching: manual and automatic.
Sensitivity, mm/mV: 5, 10, 20.
Calibration signal: automatic and manual.
Diagnostic frequency range 0.67–150 Hz or better
Filters for mains frequency, low frequency, muscle artefact, high frequency
Recorder:
Recording method: thermal paper
Recording speed, mm/sec: 25/ 50 user selectable
Channels acquired simultaneously:1/ 3 user selectable
Channels printed simultaneously:1/ 3 user selectable
Other features:
Portable
Mains and internal rechargeable battery operation
Battery operating time, minimum: 90 min.
Power requirements: 220 V, 50 Hz
Accessories:
1. Patient cable
2. 6 chest limb electrodes
3. 4 limb electrodes
4. 4 strap electrodes
5.
6. 1 bottle ECG Gel
7. 2 rolls of paper or Z-Fold
8. Carry bag
Standards
CE;
EC Marked
US FDA;
ISO certification
9
Operating theatre light (5
spotlights)
3
Surgical Lighthead with five Projectors , Ceiling Suspended
Light intensity (Ec) at least150 000 lx
Color rendering index (general) Ra 96
Color rendering index (red) R9 89
Light field size 15 – 30 cm.
Colour temperature 4200º Kelvin
Working range without focusing approximately 70 – 140 cm
Power supply 220 VAC, 50 Hz
Adjustable light intensity
Power Supply , 220V/50 Hz.
Power Stabilizer shall be supplied
This equipment shall be supplied with all necessary fixation gears to the Building
structure
Replacement Bulbs (5) and Sterilizable Handles (5) shall be supplied
Standards
CE;
EC Marked
US FDA;
ISO certification
Accessories/Spares :
5 Bulbs per unit
2 Handles per unit
10
Operating Table
(Trauma/Multifunction,
electric/hydraulic driven
Operating table designed for carrying out treatments, dressing interventions and
operations of general surgery and traumatology
All exposed metallic parts shall be made from stainless, acid proof steel.
The table base shall be mobile and shall have central brakes
Back rest and leg rest inclination angle, kidney elevator, Trendelenburg and
reverse-Trendelenburg positions and height adjustment of the table top shall be
HSMP Equipment & Furniture Component HPIU: HH
equipment specifications[1].doc
16/03/2011 page 6 of 27
1
activated by electro - hydraulic system
The table top shall be translucent for x-rays with 5 separate Sections
Table top length approximately 2000 mm
Table top width approximately 600mm
Minimal height of the table 750mm
Maximal height of the table 1100mm indicatively
Trendelenburg at least 25
Reverse Trendelenburg at least 25º
Lateral tilt at least 18º
Back rest inclination angle at least +55º to -25º
Head rest inclination angle at least 45º up
Head rest inclination angle at least 20º down
Power requirements
220 VAC ±10 % , 50Hz
Accessories:
Anaesthetic Screen with clamp with telescopic tubes : 1
Body restraint Strap with clamp : 1
Padded Shoulder supports : 2
Padded Leg support with swivel type clamp : 2
Padded arm Rests 450 -500 mm long with two arm Clamps : 2
Padded Lateral support with universal attachment Clamp : 2
Padded rubber mattresses with Anti Microbial agent incorporated into all components
that assists in Prohibiting growth of bacteria & fungi and easy to clean and maintain
and of at least of 1” thickness : 2 X5 Sections
Head rest , I.V drip Stands attachable to the table
Environmental factors:
The unit shall be capable of being stored continuously in ambient temperature of 0 -
50deg C and relative humidity of 15-90%
The unit shall be capable of operating continuously in ambient temperature of 10 -40
deg C and relative humidity of 15-90%
Accessory for traumatology are optional and shall be listed separately
Standards
CE;
EC Marked
US FDA;
ISO certification
11
Operating Table
(Multifunction,
electric/hydraulic driven)
2
Operating table for carrying out treatments, dressing interventions and operations
of general surgery (3-4 surgeries daily).
All exposed metallic parts shall be made from stainless, acid proof steel.
The table base shall be mobile and shall have central brakes
Back rest and leg rest inclination angle, Trendelenburg and reverse-
Trendelenburg positions and height adjustment of the table top shall be activated
by electro - hydraulic system
The table top shall be translucent for x-rays with 5 separate Sections
Table top length approximately 2000 mmTable top width approximately 600mm
Minimal height of the table 750mm
Maximal height of the table 1100mm indicatively
Trendelenburg at least 25º
Reverse Trendelenburg at least 25º
Lateral tilt at least 18º
Back rest inclination angle at least +55º to -25º
Head rest inclination angle at least 45º up
Head rest inclination angle at least 20º down
Power requirements
220 VAC ±10 % , 50Hz
Accessories:
Anaesthetic Screen with clamp with telescopic tubes : 1
Body restraint Strap with clamp : 1
Padded Shoulder supports : 2
Padded Leg support with swivel type clamp : 2
Padded arm Rests 450 -500 mm long with two arm Clamps : 2
Padded Lateral support with universal attachment Clamp : 2
Padded rubber mattresses with Anti Microbial agent incorporated into all components
that assists in Prohibiting growth of bacteria & fungi and easy to clean and maintain
and of at least of 1” thickness : 2 X5 Sections
Head rest , I.V drip Stands attachable to the table
Environmental factors:
HSMP Equipment & Furniture Component HPIU: HH
equipment specifications[1].doc
16/03/2011 page 7 of 27
The unit shall be capable of being stored continuously in ambient temperature of 0 -
50deg C and relative humidity of 15-90%
The unit shall be capable of operating continuously in ambient temperature of 10 -40
deg C and relative humidity of 15-90%
Standards
CE;
EC Marked
US FDA;
ISO certification
12
Anesthesia trolley
(anesthesia machine with
ventilator and monitor)
3
Anesthesia Machine used for delivering anesthesia Agents to Patient during Surgery
Procedures. The complete set-up shall also ventilate the Patient
Patient monitoring system ECG, EtCO2, Pulse Oximeter and airway pressure, NIBP,
rectal/&skin Temperature.
The full System shall be completed with :
Anaesthesia gas delivery system, O2/N2O/Compressed Air .
Circle absorber system.
Precision vaporiser for halothane
Anaesthesia ventilator
Patient Circuit
Anesthesia gas delivery system:
Shall have provision for delivery of oxygen, nitrous oxide and medical air with
pressure gauges.
Oxygen and Nitrous oxide should be linked either mechanically or pneumatically to
ensure a minimum of 25% oxygen delivery at all times to avoid delivery of hypoxic
mixture.
Shall have audio-visual oxygen Failure Warning System with Nitrous oxide cut off.
Shall have back bar ISO pin type to attach Vaporizer easily.
Flowmeter oxygen : 0÷2; 2÷10 litres/min, minimum range
Flowmeter N2O : 0÷2; 2÷10 litres/min, minimum range with safety mechanism
(hypoxic link) or similar
Flowmeter , Medical Air 0÷10 litres/min.
Shall be supplied with necessary Attachments for use of the breathing Circuits
(Ruben , Bains, Jackson-Rees or Magill )
Construction :
Shall have top Shelf to keep monitors and a Tabletop with Drawers to keep
anaesthetic drugs, equipments etc.
Castor wheels should be durable , moisture resistant and antistatic
The Anaesthesia machine frame shall be made of rust proof material/Stainless steel
or sheet Steel protected with Epoxy-Paint
Standard Circle Absorber System
Shall have adjustable pressure limiting Valve, breathing circuit pressure measuring
device.
Shall have a Bag/Ventilator selecting Valve integrated onto the Absorber.
Facility to attach oxygen sensor
Shall have CO2 absorbent Chamber Canister (Soda Lime )
Precision Vaporizers (Temperature, Pressure and Flow compensated) for Halothane
Should be easy to mount and dismount from the back bar.
Vaporisers shall have ISO pin type (Selectatec) Mounting and Vaporiser interlocking
facility
Shall have a standard filling port with keyed filling device
Should be designed for transport with liquid in Vaporiser Chamber with protection
against tipping and shaking
Maintenance free Vaporiser
Ventilator
Shall be a bag in bottle anaesthesia Ventilator with standing (ascending/Piston)
Bellows
Shall be supplied with adult and paediatric Bellows
Shall be able to set Tidal Volume, Respiratory rate and I:E ratio
Ventilator should have audible alarms for ventilator failure, low oxygen supply
Pressure, inadequate volume delivery, disconnection alarm, and power supply failure.
Ventilation mode: CMV-SV – manual , Minute volume: 2 – 30 l/Mn
Tidal volume: approx.50 – 1400ml , Breathing frequency: approx.6-60 times/min
HSMP Equipment & Furniture Component HPIU: HH
equipment specifications[1].doc
16/03/2011 page 8 of 27
Flow rate range: 5 – 60 l/min
High pressure relief valve
Power supply 220VAC, 50Hz
Accessories :
Accessories required for Ventilator :
Resuscitators , Adult / Child : 1 Each
Set of face masks , 3 Sizes : 2 Each
Corrugated hoses and tubing set for Pipeline O2 : 2 Sets
Corrugated hoses and tubing set for Pipeline Air : 2 Sets
Accessories required for Anaesthesia Machine :
Hose assembly for piped Oxygen supply : 2 Sets
Hose assembly for piped Nitrous Oxide supply : 2 Sets
Hose assembly for piped Air supply : 2 Sets
Regulator and manometer for the pipe pressure hose ( 60 psig) : 1 Set
Connection from ventilator to anaesthetic apparatus , autoclavable/reusable : 1 Set
Test bag, 1 litre : 2
Peak Flow Meter : 1
Power cord with grounding wire : 1
Dust cover : 1
Bellows assembly, adult / Children : 2 Each
Patient Circuit , Magill Type , Complete , Adult / Children . . Shall be Reusable : 2
Each
Spirometer (Peak Flow Meter for Anaesthesia procedures)
Consumables for anaesthetic unit for 2 years operation:
Battery cell (as appropriate)
Soda lime for circle absorber (5Kg/pack) : 10 Packs for 1 unit
Standards
CE;
EC Marked
US FDA;
ISO certification
13
Syringe pump
4
Syringe Driver for Parenteral Infusions , in Wards, OP, ICU's and during
Transportation/Transfer
Single Channel. Shall accept Up to 60cc Syringes , Non-Captive
Self checking of Electronics with audible call back on locked-out Controls
Tamper and fluid resistant Design . ABS Casing (Impact-Resistant )
Shall be fitted with IV Rod, Adjustable Height with 4 Hooks.Mobile on 5 Star
Stand with Castors
Internal rechargeable battery
Rate: 0.1 to 99.9 ml ( in 0.1 ml steps) indicatively
Total volume infused: 0.1 to 999.9 ml
Precision : +/-3%.Stipulated Syringes
Occlusion release shall give minimum Bolus effect
Syringes : 10 Types at least of LuerLock Type 10/60mL.Syringes,+ Specific
Choice
Pause infusion facility
Alarms/Safety
Infusion nearly complete
Infusion complete
Occlusion
System malfunction
Syringe unlocked
Plunger disengaged
Low battery
Protection against Leakage Current:Type CF.Equipment
Protection against Electric Shocks:Class II.Equipment
Safety features: Error coding (event log)
Rate change through Audible Stop Switch
Audible Call back on locked-Out-Controls
Automatic detection of Syringe Volume
Minimum bolus on Occlusion Release at any selected Rate
Communication : RS232C Interface
Electrical:
Electrical: Class II (double insulated)
HSMP Equipment & Furniture Component HPIU: HH
equipment specifications[1].doc 16/03/2011 page 9 of 27
Bat.Autonomy : 8 Hrs.Minimum
Power supply: 220 VAC.50.60Hz
External transformer . Stabilizer
Environmental factors
Shall meet IEC-60601-1-2: 2001(Or Equivalent BIS)
General Requirements of Safety for Electromagnetic Compatibility or should
comply with 89/366/EEC; EMCdirective.
The unit shall be capable of being stored continuously in ambient
temperature of 0 -50deg C and relative humidity of 15-90%
The unit shall be capable of operating continuously in ambient temperature of
10 -40deg C and relative humidity of 15-90%
Accessories
Coupler between 2 units
Mobile floor stand
IV stand clamp
Rail clamp
Standards :
IEC601.1/IP33
Safety: IEC 601 / 1. CF.
Compliance with EN 60601.1 and PrEN 60601-1-24.
CE 0459 marking in compliance with EEC 93/42 Medical Products directive.
Comply with EN60601-1-2, EN60601-2-24
EN ISO 9001:2000, EN ISO 13485:2000; EN 46001:1996
14
Electrosurgical
Unit (Monopolar-bipolar)
3
HF electrosurgical unit shall be used to execute monopolar and bipolar surgery in
many fields of application where high precision and reliability are essential
Outputs of cut, coagulate and blend
Maximum output 300 W for monopolar cut
Activation :
Double pedal switch which may be used for the monopolar and bipolar functions .
Hand-switch handle
Bipolar electrode with pedal switch or with automatic Start/Stop system ( for
coagulation only)
Control
The Unit shall stop automatically in case of internal error which shall be identified on
Display and with audible alarm
Memorization : User shall be able to use at least 4 working programs
Safety :
Neutral plate safety circuit shall control connections and contacts of Neutral Plate
with Tissues:
Defective Contact shall be notified with visual Alarm and immediate reducing of
power
Output circuit : floating - protected against defibrillator interferences . Shall have HF
leakages less than 150mA through each electrode
Power Supply : 220VAC, 50Hz
Cooling: convection without fan
Accessories :
Single-use two button Handle , Autoclavable
Handle shall be provided with a 3 Pins socket that shall fit majority of bipolar
electrosurgical units
Operative Foot-switch (usable as alternative to handle)
Reusable neutral Plate
Kit of 10 short autoclavable Electrodes
Electrode wire - straight
Electrode wire - angled 45°
Electrode bend - 0 4 - straight
Electrode bend - 0 8 - straight
Electrode hook - angled 45°
Electrode wire 1 mm.- angled 45°
Electrode straight slip-knot - straight
Electrode angled slip-knot - straight
Electrode ball point - 0 3 mm. - angled 45°
Electrode ball point - 0 3 mm. – straight
Bipolar Forceps : 6
Forceps shall be provided with a standard European connection
Straight Forceps - 18 cm. (7")
HSMP Equipment & Furniture Component HPIU: HH
equipment specifications[1].doc 16/03/2011 page 10 of 27
Curved Forceps - 18 cm. (7")
Curved Forceps - 20 cm. (7 3/4")
Bayonet Forceps - 18 cm. (7")
Bayonet Forceps - 20 cm. (7 3/4")
Straight Forceps - 20 cm. (7 3/4")
Cable, Bipolar
Adaptor, Bipolar Cable
User Manual
Standards
CE;
EC Marked
US FDA;
ISO certification
15 Suction Pump
2
Suction Unit for Major Surgery Procedures . Mains-powered , mobile on 4 antistatic
Castors, ABS Casing and 2 graduated Canisters of 2,000ml each made of
Polycarbonate autoclavable at 121°C and disposable suction bags
Shall require no maintenance nor lubrication
Oil-free pump, maximum suction of at least 500 mm Hg
Free flow rate at least 25 l/min
Main Switch with Pilot Lamp . Fuses
Pedal Action
Shall be equipped with a protective thermal cut-out relay.
Shall be equipped with motor-protection cap that totally prevents aspirated liquids or
secretions from reaching and damaging the vacuum pump
Suction command with continuous adjustment , Vacuometer
2X2,000ml Canisters with airproof screwing-cap with independent Overflow devices
. Fast Connectors and silicone Tubing
Power Supply : 220VAC.50Hz.
Ventilation Fan for overheating
Sound level: Shall be not more than 55 dBA
Accessories
Silicone Tubing , sterilizable
Transparent Cannula Holder, Sterilizable
Anti-Bacterial Filters ( 4 )
Set of 4 canulaes with Holder : Yankhauer , Soft Universal Yankhauer
Diameter : 8.0/6.0mm with anti-sticking Lumen and High Suction Lumen
Universal Soft Canulaes diameter : 6.0/4.0mm
Frazier Canulaes (Fergusson) diameter :1,5/2.0/3.0/4.0mm
Jackson Canulaes : 35x3 /45X3/25X4/35X4/45X4/55X4/60X4cm
Standards
CE;
EC Marked
US FDA;
ISO certification
16 Operating instrument set
2
1 Operating Scissors, 170 mm, straight, blunt-blunt points 1
2 Operating Scissors, 140 mm, straight, sharp-blunt points 1
3 Operating Scissors, 140 mm, curved, sharp-blunt points 2
4 Dissecting Scissors, curved, for deep surgery, 230 mm 2
5 Operating Scissors, 140 mm, curved, blunt-blunt points 4
6 Operating Scissors, curved, blunt-blunt points, 170 mm 2
7
Operating Scissors, 140 mm, curved, hardened, blunt-blunt
points
1
8 Sharp Tips Scissors, curved, 160 mm 1
9 Dressing Forceps, standard pattern __ 150_2,5 6
10 Dressing Forceps, standard pattern __ 200_2,5 4
11 Surgical Forceps, 150x2,5 8
12 Surgical Forceps, 1 x 2, 200x2.5 mm 4
13 Abdominal Depressor, 285 mm, blade 70 mm wide 1
14 Tissue Forceps, 265 mm, straight 5
15 Tissue Forceps, 256 mm, curved 5
HSMP Equipment & Furniture Component HPIU: HH
equipment specifications[1].doc 16/03/2011 page 11 of 27
16 Hemostatic Forceps, curved, atraumatic (160 mm) 6
17
Hemostatic Forceps, curved, 1x2 teeth, serrated, size 2,
(160mm)
6
18 Hemostatic Forceps, curved, serrated, size 2 , (196 mm) 6
19 Hemostatic Forceps, curved, serrated, (240 mm) 6
22
Hemostatic Forceps, straight, 1x2 teeth, serrated, size 1 ,
(150 mm)
3
23
Hemostatic Forceps, curved, 1x2 teeth, serrated, size 1,
(150mm)
6
24
Hemostatic Forceps, curved, 1x2 teeth, serrated, size 1,
(200mm)
6
25
Hemostatic Forceps for deep surgery, child size, size 2,
(225mm)
6
26 Rake Retractor, 2 sharp prongs, (200 mm) 2
27 Rake Retractor, 3 sharp prongs, (200 mm) 2
28 Rake Retractor, 200 mm, 4 sharp prongs, style 2 2
29 Rake Retractor, 2 blunt prongs, 200 mm 3
30 Rake Retractor, 220 mm, 4 blunt prongs, style 2 4
31 Rake Retractor, 6 sharp prongs, (200 mm) 2
32 Abdominal Retractor 2
33 Needle Holder,( 200 mm) 2
34 Needle Holder, ( 250 mm) 2
35 Vascular Needle Holder , 160 mm 2
36 Vascular Needle Holder, 200 mm 2
37 Intestinal Forceps, non-crushing, curved, child size, (192mm) 2
38 Intestinal Forceps, non-crushing, curved, adult size, (235mm) 2
39 Tongue Spatula, straight, 20_16mm, (180 mm) 2
40
M_SQUITO Hemostatic Forceps, serrated, curved on edge,
61/10" (154 mm)
6
41
Intestinal Forceps, non-crushing, straignt, adult size,
(240mm)
2
42 Abdominal Retractor (60 mm) 2
43 Abdominal Retractor (100 mm) 2
44 Liver Speculum (100 mm)
2
17 Suction Pump
1
Suction Unit for Emergency procedures , fitted with single 1000ml capacity
Polycarbonate Canister , mobile on 4 Castors and carrying handle
The Unit shall enable operation from either mains supply, or independently from an
internal 12V rechargeable battery
The unit shall recharge when plugged into the mains
An indicator on the casing shall indicate the status of the battery . A Vacuometer
shall indicate the suction value
Maintenance free diaphragm pump
Overflow safety device
The Unit shall be capable to generate 500mm Hg at least
Free flow rate at least 25 litres per min
Shall be fitted with easily replaceable anti-bacterial filter with 5 spare
Shall provide over 30 minutes of full vacuum from a fully charged battery
Fitted with rechargeable battery
Operating mains voltage: 220 V - 50 Hz
Recharging is by trickle charge and shall take eight hours maximum to restore full
charge
Accessories :
Hand piece with regulator and appropriate probes shall be supplied as well as an
additional Canister
Standards
CE;
EC Marked
US FDA;
ISO certification
18 Defibrillator
2
The following specifications define a portable, lightweight ; battery or accumulator
operated automated external to treat Patients requiring basic Life Support
Weight of complete unit, excluding batteries or accumulator , shall permit easy
Transport . The overall Dimensions shall enable easy storage and portability
HSMP Equipment & Furniture Component HPIU: HH
equipment specifications[1].doc 16/03/2011 page 12 of 27
The Casing shall be constructed to withstand the standard operating Conditions
in a Physician’s office and/or Hospital. The device must also with stand the
harsh operating conditions associated with Ambulance use. The unit shall be safe to
use both for the operator and the patient.
Defibrillator
The device shall utilize Biphasic waveform technology as the primary energy delivery
option and shall include automatic and manual modes
The biphasic waveform shall allow for escalating energies up to at least 200 Joules
Safety: Internal discharge of energy if no discharge is effected.
Internal self-testing and fault recognition.
Required power 200-240V 50/60 Hz
Weight not more 10kg
Accessories :
Full Standard Accessories Set-Up shall be supplied
Standards
CE;
EC Marked
US FDA;
ISO certification
19 Ventilator ICU
2
Intensive care ventilator, for use on Pediatric/Infant- Adult patients
With Graphics Display Monitor.
Breath Types: VC, PC, VTPC
Modes ( in each breath Type): A/CMV, SIMV, SPONT,
Spontaneous Breath Choices: PS, VTPC.
Features:
Inspiratory time at least 0-3 sec with optional pause
Inspiratory flow at least 3-140 l/min
Pressure support: at least 0 to 45cm H2O/mbar
Tidal Volume: at least 10-800ml
Resp.Rate: at least 6-150b/min
Pressure Limit: 0 to 80cmH2O/mbar
I:E Rate: 1:4 to 4:1
Trigger pressure or flow
FiO2: 0.21 to 1.00
PEEP/CPAP: 0 to 45 cmH2O/mbar
Leak Compensation
Standby Condition: Allows setting to be preset and Circuit Check tests to be
performed prior to starting Ventilation
Open Exhalation Valve (BPRV): On/Off for Biphasic Pressure Release
Ventilation
Event History Log: Records 1000 events; alarm & settings
Compliance Compensation: On/Off
Graphical display of respiratory parameters
Must be mounted over the Mobile Medical Air Compressor.
Power requirements
220 VAC ±10%, 50Hz.
Internal rechargeable Battery provides an minimum. 60 minutes for complete
ventilator function.
Accessories :
Medical Air Compressor
Humidifier
1 Tube set, adults
1 Tube set, Children
1 Humidifier chamber
O2-connecting Tube 5.0 m
1 Air-connecting Tube 5.0 m
Humidifier : Heating power with safety thermostat and indication of operating status
1 Circuit , Adults with patient Tubes , water traps, Y-Piece,mask elbow, catheter
connector
1 Circuit , Children with patient Tubes, water traps, Y-Piece,mask elbow, catheter
connector
1 Carriage
Articulated Bracket for rail
O2-cell without housing
HSMP Equipment & Furniture Component HPIU: HH
equipment specifications[1].doc 16/03/2011 page 13 of 27
Sensors 5 pcs set
Temperature Sensor
Humidifier
Bag- test-lung
Standards
CE;
EC Marked
US FDA;
ISO certification
20 Pulsoximeter
2
Real-time SpO2%
Oxygen Saturation Range (%SpO2)
Measurement range at least 70-99%
Accuracy (%SpO2) ±2% at100%~80%, ±3% at 80%~70%
Real-time heart rate (BPM)
Pulse
Measurement range 30-250 bpm
Accuracy ± 1 bpm at 30~250 bpm
Display
SpO2,Pulse rate, Pulse signal wave, Max/Min setup (SpO2% / BPM)
Power Requirements internal rechargeable battery providing an operating time of at
least 8 hours
Compact design for easy transportability
Accessories:
3 rolls of paper
Operation manual
1 adult reusable finger Sensor
1 Neonate Sensor
Standards
CE;
EC Marked
US FDA;
ISO certification
21 Patient Monitor, ICU
7
Patient monitor that shall draw Cardiograms, measures Blood Pressure (invasive or
not-invasive type), Respiration values, Temperatures at two spots, Oxygen Saturation
in arterial blood ( SpO2 ), and Pulse rate ( HR ) .
May be used to Monitor Patients on the transportation by using batteries, and the
measured information on patient's condition shall be displayed in wave forms along
with numerical values.
Display/Screen
Display : Colour TFT display , screen diagonal Curves and measured values : not less
than 25.00cm . Up to 6 Curves as well as numerical fields
Brightness: can be manually adjusted
Sweep Speed : Shall adjuste to 6.25, 12.5, 25 and 50 mm/s
Display pixels: at least 1024x768
Monitor Interfaces
R-SYNC: for defibrillator , Each R-wave shall be clearly identified , Staff call:
Remote alarm via alarm adapter
RS-232: as computer interface
Mains/Battery Operation
Mains operation: shall be via built-in Power Supply Battery operation: via integrated
battery or Charging console . Battery operating time: approx. up to 4 hrs at least
Power Supply : 220 V AC ±10%, automatic change-over Power failure buffer : for
trend values,
System Expansion
Software updates via interfaces
Monitoring parameters shall include : ECG, RESP, IBP, NIBP, SPO2, PULSE RATE
, DUAL-TEMP
S-T segment synchronized detection and analysis
Graphic and tabular trends of all parameters for at least 96 hours
20-Mn minimum ECG real time waveform and parameters storage and review
Shall be able to be connected with a Central Monitor Station
HSMP Equipment & Furniture Component HPIU: HH
equipment specifications[1].doc 16/03/2011 page 14 of 27
Inner high-definition thermal dot matrix recorder optional, which can output
waveforms and characters
ECG : Input : Whole-lead ECG cable, standard AAMI cable connector
Sweep speed: 12.5, 25, 50mm/s 0.5~40Hz(for monitor or operation)
Calibration signal: 1mV , Protection: Against electrosurgical interference and
defibrillation
Heart Rate: Measuring range: 30~250bpm , Alarm mode: Audible and
visualAlarm
Respiraton Measurement Range: at least 0~60 bpm
Oxygen Saturation(SPO2) Meas. Range: at least 70~100%
NIBP : Method: Oscillometric , Mode: Manual, Auto , Resolution: 1 mmHg
Shall have Over-pressure protection: (Adult 300mmHg ; Child 220mmHg
Temperature : Meas. Range: at least 20~50C , Resolution: 0.1 C , Accuracy: +/- 0.2 C
Operation Environment : Temperature: 0~45C , Humidity: 30~85%(non-condensing)
Accessories
Electrode cable for disposable electrode (3 leads)
NIBP cuff for adults
NIBP cuff for children
Hose for NIBP cuff
Respiration Sensor
Power cord
Disposable electrode (5 pcs)
Temperature sensor - Adult (rectal)
SpO2 sensor reusable
Console / Bedside Holding Device
Standards
CE;
EC Marked
US FDA;
ISO certification
22 Patient Monitor, anesthesia
2
Patient monitor that shall draw Cardiograms, measures Blood Pressure (invasive or
not-invasive type), Respiration values, Temperatures at two spots, Oxygen Saturation
in arterial blood ( SpO2 ), and Pulse rate ( HR ) , End Tidal CO2 (etCO2) .
May be used to Monitor Patients on the transportation by using batteries, and the
measured information on patient's condition shall be displayed in wave forms along
with numerical values.
Display/Screen
Display : Colour TFT display , screen diagonal Curves and measured values : not less
than 25.00cm . Up to 6 Curves as well as numerical fields
Brightness: can be manually adjusted
Sweep Speed : Shall adjuste to 6.25, 12.5, 25 and 50 mm/s
Display pixels: at least 1024x768
Monitor Interfaces
R-SYNC: for defibrillator , Each R-wave shall be clearly identified , Staff call:
Remote alarm via alarm adapter
RS-232: as computer interface
Mains/Battery Operation
Mains operation: shall be via built-in Power Supply Battery operation: via integrated
battery
or Charging console . Battery operating time: approx. up to 4 hrs at least
Power Supply : 230 V AC ±10%, automatic change-over Power failure buffer : for
trend values,
System Expansion
Software updates via interfaces
Monitoring parameters shall include : ECG, RESP, NIBP, SPO2, PULSE RATE ,
DUAL-TEMP and etCO2
S-T segment synchronized detection and analysis
Graphic and tabular trends of all parameters for at least 96 hours
20-Mn minimum ECG real time waveform and parameters storage and review
HSMP Equipment & Furniture Component HPIU: HH
equipment specifications[1].doc 16/03/2011 page 15 of 27
Shall be able to be connected with a Central Monitor Station
Inner high-definition thermal dot matrix recorder optional, which can output
waveforms and characters
ECG : Input : Whole-lead ECG cable, standard AAMI cable connector
Sweep speed: 12.5, 25, 50mm/s 0.5~40Hz(for monitor or operation)
Calibration signal: 1mV , Protection: Against electrosurgical interference and
defibrillation
Heart Rate: Measuring range: 30~250bpm , Alarm mode: Audible and visual Alarm
Measurement Range: at least 0~60 bpm
Oxygen Saturation(SPO2) Meas. Range: at least 70~100%
NIBP : Method: Oscillometric , Mode: Manual, Auto , Resolution: 1 mmHg
Shall have Over-pressure protection: (Adult 300mmHg ; Child 220mmHg
Temperature : Meas. Range: at least 20~50C , Resolution: 0.1 C , Accuracy: +/- 0.2 C
End Tidal CO2 (etCO2) Module , Non Dispersive Infrared Spectroscopy Method
Mainstream or sidestream
Displayed Data : Waveform labels and annotations , etCO2, inC02 and respiration
rate values
etCO2 Alarm Limits
Operation Environment : Temperature: 0~45C , Humidity: 30~85%(non-condensing)
Accessories
Electrode cable for disposable electrode (3 leads)
NIBP cuff for adults
NIBP cuff for children
Hose for NIBP cuff
Respiration Sensor
Power cord
Disposable electrode (5 pcs)
Temperature sensor - Adult (rectal)
SpO2 sensor reusable
EtCO2 Mainstream or Sidestream accessories , Transducer , Lines , Filters
Console / Bedside Holding Device
Standards
Classification: CF, defibrillation-proof,
CE;
EC Marked
US FDA;
ISO certification
23 Oxygen Concentrator
1
Compact and easy to transport oxygen concentrator ( Mobile on Castors )
The Oxygen Concentrator shall feature a dual-head Compressor. One head
producing. Capacity : 1 to 5 l/Mn of O2 at 90% minimum at maximum flow
Pressure-compensated Flowmeter shall permit use of long canula while maintaining
accurate flow setting.
Safety Alarms: Audible and visual : Power Failure , Restricted Flow , , Low O2
Pressure-relief Valve and thermal protection of the Compressor
Double-insulated Unit , Two-prong plug
Flame-retardant Cabinet
Sound Level (ANSI): 50 dB average
Fixed humidifer Port and Recess shall prevent bottle and connector breakage
Power Requirements: 220 VAC,50 Hz
Accessories :
O2 Tubing , 3 Face Masks ( Adult , Infant , New Born ), Humidifier ,
Set of Filters
Standards
CE;
EC Marked
US FDA;
HSMP Equipment & Furniture Component HPIU: HH
equipment specifications[1].doc 16/03/2011 page 16 of 27
ISO certification
24 Couch, Gynecology
4
Manufactured from chrome plated steel tube, three sections with independent
manual controls and Trendelenburg position.
Top is upholstered with washable plastic material and flame retardant.
Complete with padded Goepel leg-holders and S.S. bowl.
Knock down construction.
25 Delivery Bed
2
Manufactured from reinforced epoxy-painted steel tube
Three sections with independent manual controls.
Trendelenburg position.
Third section extensible.
Top is upholstered with washable plastic material flame retardant.
Complete with S.S. tray, and two adjustable knee crutches, padded shoulder rest,
handgrips, leg-holders.
26 Examination Lamp
3
Lamp , Examination . Mobile on 5 Castors , 5 star-shaped Heavy and stable Stand .
Halogen Type
Height of LampHead shall be adjustable
Light Intensity shall be at least 15000 Lux with a Field radius of 17cm Minimum
Color Intensity : 4300 °K. Single Reflector of 150mm Minimum
Power rating : 1X40W Minimum . 220VAC/50Hz . Transformer , 24V DC shall be
included
Accessories /Spares : 5 Halogen Bulbs
Standards
CE;
EC Marked
US FDA;
ISO certification
27 Uterine Aspiration Set
1
Stainless steel with handle for suction regulation.
Cannule ø mm. 6/8/10/12/14
28 Fetal Monitor
1
CTG or Fetal monitor for recording and analyzing the fetal heart rate (FHR) on beatto-
beat basis and the uterine contraction with Doppler -US transducer and taco
transducer.
Portable handheld CTG monitor. Compact and battery operated, shall allow Patients
to be continuously monitored anywhere . Color display shall show the CTG data
presented in standard format for easy interpretation.
Fetal signals shall be easily located using the wide angle transducers with
contractions stress being monitored by an external pressure transducer. Audio signals
of each fetal heart are available on selection by the user through the built in Audio
system.
The user will have control over the audio volume, audio fetal channel, chart recorder
speed, toco zero, and event mark via the units front panel push buttons.
Directional and watertight Ultrasound and TocoGraph with color-coded transducers
for Single and Twins
Transducer: Multi element wide angle
Range: 30 to 240 bpm
Power output: < 20mW/cm2
Toco
Transducer: Real Time uterine Activity (contractions) trace on standard scale format
Digital autocorrelation shall provide accurate and reliable FHR traces
A pre-cut, non-fray elasticised belt with Buckle shall enable easy Transducer
positioning for more accurate traces in difficult monitoring situations
Clinical Event marker & automated foetal movement detection . Trace annotation
Comprehensive patient database
Shall store over 500 patient records (30 minutes traces) in internal memory of FHR,
Toco , Foetal Movements with unlimited data storage on plug-in PCMCIA memory
Cards
Computer interface 3 wire RS232
Thermal printer for common CTG paper
Speeds: 1,2,3, cm/min
Paper: z-fold pre-printed chart scale
The grid and the FHR/UA trace shall be printed simultaneously , ensuring printing
accuracy and preventing potential errors in interpretation due to paper drifting or pen
HSMP Equipment & Furniture Component HPIU: HH
equipment specifications[1].doc 16/03/2011 page 17 of 27
jitter
Power Supply :220 V AC and DC battery operated
Accessories:
1. Belts : 1 Original + 1 Spare Set
2. Buckles
3. IUP Starter Kit
4. Multicrystal transducer
5. Toco Transducers
6. Event marker Consumables:
7. CTG printer paper : 20 Packs
8. Fetal scalp electrodes
9. Ultrasound gel
10. IUP catheters
Standards
CE;
EC Marked
US FDA;
ISO certification
29 Scale for newborn children
3
Capacity-0-15 kg
Graduation-10 gr.
Precision-+/- 10 gr.
Washable surface
30 Microscope Binocular
1
Body with 6V 20W Halogen for 220V, 50 Hz.
Quadruple revolving nosepiece.
Sturdy,base with supportive rubber feet.
With low position, coaxial coarse and calibrated fine focus control, with graduation
reading 2 microns per division.
Total focusing 40 mm.
Condenser adjustable.
Built-in Power supply , halogen bulb adjustable.
Double layer mechanical specimen stage.
Dust cover, blue filter, immersion oil and instruction manual included.
2 pcs. 6V 20W halogen bulbs included.
Binocular Head, 30° inclined, 360° rotatable, Interpupillary Distance 55 – 75 mm
Objectives:
ACHRO 4x/0.10
ACHRO 10x/0.2
ACHRO 40x/0.65, spring loaded
ACHRO 100x/1.25, Oil, spring loaded
31 Microscope Monocular
1
Microscope with metal base with 360° rotatable monocular tube, 45° inclined
3-fold objective nosepiece
separated coarse and fine focus
Stage focus control (protection of sample!)
Objectives:
ACHRO 4x/0.10
ACHRO 10x/0.25
ACHRO 40x/0.65 spring
Objective 100x/1.25 oil,
attachable mechanical stage, etc.
Abbé condenser n.A. 1.25 with iris diaphragm, variable at height
Stage with stage clips
Blue filter.
Built-in illumination 6V/20 Watt Halogen, adjustable
Daylight mirror for outdoor-use
Dust cover.
Packed in styrofoam case
Power Requirements: 220V / 50 Hz.
HSMP Equipment & Furniture Component HPIU: HH
equipment specifications[1].doc 16/03/2011 page 18 of 27
32 Analyzer Hematology
1
To provide accurate and precise hematology results with WBC 3 parts differential.
Principle method of detection: volumetric impedance, electrical impedance, laser
flow cytometry, hydrodynamic focusing or equivalent.
It should be able to provide precise measurement of
1) White Blood Cells count (WBC): Precision (CV, coefficient of variation) < 3.5%
2) Red Blood Cells (RBC) : Precision (CV) < 2%
3) Platelets (PLT) : Precision (CV) <6 data-blogger-escaped--="" data-blogger-escaped-.0="" data-blogger-escaped-.="" data-blogger-escaped-0-120="" data-blogger-escaped-0-250="" data-blogger-escaped-0-="" data-blogger-escaped-0.1="" data-blogger-escaped-0.5="" data-blogger-escaped-10-40="" data-blogger-escaped-1000="" data-blogger-escaped-100="" data-blogger-escaped-100l="" data-blogger-escaped-10="" data-blogger-escaped-11="" data-blogger-escaped-12="" data-blogger-escaped-134="" data-blogger-escaped-13="" data-blogger-escaped-14="" data-blogger-escaped-15-90="" data-blogger-escaped-150="" data-blogger-escaped-15="" data-blogger-escaped-16="" data-blogger-escaped-18="" data-blogger-escaped-19="" data-blogger-escaped-1="" data-blogger-escaped-2001="" data-blogger-escaped-20="" data-blogger-escaped-21="" data-blogger-escaped-220-240vac="" data-blogger-escaped-220="" data-blogger-escaped-220v="" data-blogger-escaped-220vac.50hz.="" data-blogger-escaped-220vac.50hz="" data-blogger-escaped-220vac="" data-blogger-escaped-232="" data-blogger-escaped-240vac="" data-blogger-escaped-24="" data-blogger-escaped-250="" data-blogger-escaped-25="" data-blogger-escaped-27="" data-blogger-escaped-2="" data-blogger-escaped-3.0="" data-blogger-escaped-300="" data-blogger-escaped-30="" data-blogger-escaped-33="" data-blogger-escaped-34="" data-blogger-escaped-35="" data-blogger-escaped-36="" data-blogger-escaped-37="" data-blogger-escaped-38="" data-blogger-escaped-39="" data-blogger-escaped-3="" data-blogger-escaped-4.="" data-blogger-escaped-40="" data-blogger-escaped-41="" data-blogger-escaped-42="" data-blogger-escaped-43="" data-blogger-escaped-44="" data-blogger-escaped-45="" data-blogger-escaped-4="" data-blogger-escaped-50-="" data-blogger-escaped-5000="" data-blogger-escaped-50="" data-blogger-escaped-50hz.="" data-blogger-escaped-50hz="" data-blogger-escaped-5="" data-blogger-escaped-60="" data-blogger-escaped-60l.="" data-blogger-escaped-6850.1987="" data-blogger-escaped-6="" data-blogger-escaped-7="" data-blogger-escaped-80="" data-blogger-escaped-8="" data-blogger-escaped-950x109="" data-blogger-escaped-99="" data-blogger-escaped-9="" data-blogger-escaped-:="" data-blogger-escaped-:no.="" data-blogger-escaped-a="" data-blogger-escaped-abnormal="" data-blogger-escaped-above="" data-blogger-escaped-accept="" data-blogger-escaped-accessories:="" data-blogger-escaped-accessories="" data-blogger-escaped-accuracy="" data-blogger-escaped-acid="" data-blogger-escaped-action="" data-blogger-escaped-adapted="" data-blogger-escaped-adaptor="" data-blogger-escaped-addition="" data-blogger-escaped-adequately="" data-blogger-escaped-adjustable="" data-blogger-escaped-adult="" data-blogger-escaped-adults="" data-blogger-escaped-air="" data-blogger-escaped-alarm="" data-blogger-escaped-albumin="" data-blogger-escaped-all="" data-blogger-escaped-alp="" data-blogger-escaped-also="" data-blogger-escaped-alt="" data-blogger-escaped-ambient="" data-blogger-escaped-ambu="" data-blogger-escaped-an="" data-blogger-escaped-analogical="" data-blogger-escaped-analysing="" data-blogger-escaped-analysis.="" data-blogger-escaped-analysis="" data-blogger-escaped-analyzer="" data-blogger-escaped-and="" data-blogger-escaped-anticoagulant="" data-blogger-escaped-any="" data-blogger-escaped-appropriate="" data-blogger-escaped-approximately="" data-blogger-escaped-aptt="" data-blogger-escaped-ards="" data-blogger-escaped-are="" data-blogger-escaped-as="" data-blogger-escaped-ast="" data-blogger-escaped-at="" data-blogger-escaped-atmosphere="" data-blogger-escaped-audible="" data-blogger-escaped-autoclavable="" data-blogger-escaped-autoclave="" data-blogger-escaped-automated="" data-blogger-escaped-automatic="" data-blogger-escaped-avoiding="" data-blogger-escaped-bacterial="" data-blogger-escaped-bacteriological="" data-blogger-escaped-bag="" data-blogger-escaped-balloon="" data-blogger-escaped-bar-code="" data-blogger-escaped-based="" data-blogger-escaped-basic="" data-blogger-escaped-bath="" data-blogger-escaped-be="" data-blogger-escaped-beams="" data-blogger-escaped-being="" data-blogger-escaped-benchtop="" data-blogger-escaped-between="" data-blogger-escaped-bilirubin="" data-blogger-escaped-bis="" data-blogger-escaped-blanks="" data-blogger-escaped-blood="" data-blogger-escaped-body="" data-blogger-escaped-both="" data-blogger-escaped-brakes="" data-blogger-escaped-brushless="" data-blogger-escaped-bs="" data-blogger-escaped-buckets="" data-blogger-escaped-build-in="" data-blogger-escaped-built-in="" data-blogger-escaped-bun="" data-blogger-escaped-burnout="" data-blogger-escaped-by="" data-blogger-escaped-c.="" data-blogger-escaped-c="" data-blogger-escaped-ca="" data-blogger-escaped-calculation="" data-blogger-escaped-calibration="" data-blogger-escaped-capabilities="" data-blogger-escaped-capability="" data-blogger-escaped-capable="" data-blogger-escaped-capacities="" data-blogger-escaped-capacity-5-150kg.="" data-blogger-escaped-capacity:="" data-blogger-escaped-capacity="" data-blogger-escaped-capillary="" data-blogger-escaped-carrying="" data-blogger-escaped-case="" data-blogger-escaped-category="" data-blogger-escaped-ce="" data-blogger-escaped-cell="" data-blogger-escaped-centrifuge="" data-blogger-escaped-centrifuges="" data-blogger-escaped-certification="" data-blogger-escaped-chamber="" data-blogger-escaped-changes="" data-blogger-escaped-channels="" data-blogger-escaped-chemistry="" data-blogger-escaped-children="" data-blogger-escaped-cholesterol="" data-blogger-escaped-citrate="" data-blogger-escaped-ck="" data-blogger-escaped-cl-="" data-blogger-escaped-cleaning="" data-blogger-escaped-climate="" data-blogger-escaped-closure="" data-blogger-escaped-clots="" data-blogger-escaped-cm="" data-blogger-escaped-co2="" data-blogger-escaped-coagulometer="" data-blogger-escaped-coefficient="" data-blogger-escaped-cofactors="" data-blogger-escaped-colour="" data-blogger-escaped-compact="" data-blogger-escaped-compatibility.="" data-blogger-escaped-compensation="" data-blogger-escaped-complement="" data-blogger-escaped-complete="" data-blogger-escaped-compliance="" data-blogger-escaped-comply="" data-blogger-escaped-component="" data-blogger-escaped-components="" data-blogger-escaped-computer="" data-blogger-escaped-concentration="" data-blogger-escaped-concern="" data-blogger-escaped-conditions="" data-blogger-escaped-configuration="" data-blogger-escaped-connection="" data-blogger-escaped-consist="" data-blogger-escaped-consisting="" data-blogger-escaped-constructed="" data-blogger-escaped-consumables:="" data-blogger-escaped-consumables="" data-blogger-escaped-continuous="" data-blogger-escaped-continuously="" data-blogger-escaped-control="" data-blogger-escaped-controlled="" data-blogger-escaped-corpuscular="" data-blogger-escaped-corrosion="" data-blogger-escaped-counteract="" data-blogger-escaped-country.="" data-blogger-escaped-creatinine="" data-blogger-escaped-csf="" data-blogger-escaped-cssd="" data-blogger-escaped-curves="" data-blogger-escaped-cut-off="" data-blogger-escaped-cut-out="" data-blogger-escaped-cuvettes="" data-blogger-escaped-cyanide="" data-blogger-escaped-cycle="" data-blogger-escaped-data="" data-blogger-escaped-declare="" data-blogger-escaped-defrosting="" data-blogger-escaped-deg="" data-blogger-escaped-degree="" data-blogger-escaped-delivery="" data-blogger-escaped-designed="" data-blogger-escaped-detail="" data-blogger-escaped-detect="" data-blogger-escaped-detection="" data-blogger-escaped-determination="" data-blogger-escaped-device="" data-blogger-escaped-dial.="" data-blogger-escaped-different="" data-blogger-escaped-differential="" data-blogger-escaped-digital="" data-blogger-escaped-direct="" data-blogger-escaped-display.="" data-blogger-escaped-display="" data-blogger-escaped-distribution="" data-blogger-escaped-doc="" data-blogger-escaped-door="" data-blogger-escaped-double="" data-blogger-escaped-drainage="" data-blogger-escaped-drive="" data-blogger-escaped-dual="" data-blogger-escaped-duplicate="" data-blogger-escaped-during="" data-blogger-escaped-dynamic="" data-blogger-escaped-ec="" data-blogger-escaped-edta="" data-blogger-escaped-efficient="" data-blogger-escaped-electromagnetic="" data-blogger-escaped-electronic="" data-blogger-escaped-electronically-operated="" data-blogger-escaped-elements="" data-blogger-escaped-emergencies="" data-blogger-escaped-emergency="" data-blogger-escaped-end="" data-blogger-escaped-enhanced="" data-blogger-escaped-environmental="" data-blogger-escaped-equiped="" data-blogger-escaped-equipment="" data-blogger-escaped-equivalent="" data-blogger-escaped-etc.="" data-blogger-escaped-exact="" data-blogger-escaped-external="" data-blogger-escaped-fabricated="" data-blogger-escaped-face="" data-blogger-escaped-factors="" data-blogger-escaped-failure="" data-blogger-escaped-false="" data-blogger-escaped-far="" data-blogger-escaped-fda="" data-blogger-escaped-feel="" data-blogger-escaped-fibrinogen="" data-blogger-escaped-filters="" data-blogger-escaped-fingertips="" data-blogger-escaped-first="" data-blogger-escaped-fitted="" data-blogger-escaped-flat="" data-blogger-escaped-flow="" data-blogger-escaped-following="" data-blogger-escaped-for="" data-blogger-escaped-forced="" data-blogger-escaped-forward="" data-blogger-escaped-free="" data-blogger-escaped-freezer="" data-blogger-escaped-from="" data-blogger-escaped-front-loading="" data-blogger-escaped-front="" data-blogger-escaped-fully="" data-blogger-escaped-function="" data-blogger-escaped-furniture="" data-blogger-escaped-fuses.="" data-blogger-escaped-g="" data-blogger-escaped-gasket="" data-blogger-escaped-gb="" data-blogger-escaped-general="" data-blogger-escaped-generator="" data-blogger-escaped-generic="" data-blogger-escaped-glass="" data-blogger-escaped-glucose="" data-blogger-escaped-gr.="" data-blogger-escaped-grade="" data-blogger-escaped-grades="" data-blogger-escaped-graduation-100gr.="" data-blogger-escaped-granulocyte="" data-blogger-escaped-h2o="" data-blogger-escaped-h="" data-blogger-escaped-have="" data-blogger-escaped-heating="" data-blogger-escaped-hematocrit="" data-blogger-escaped-hemoglobin="" data-blogger-escaped-heparin="" data-blogger-escaped-here="" data-blogger-escaped-hh="" data-blogger-escaped-high="" data-blogger-escaped-hinged="" data-blogger-escaped-histograms="" data-blogger-escaped-holders="" data-blogger-escaped-horizontal="" data-blogger-escaped-hot="" data-blogger-escaped-hour="" data-blogger-escaped-hpiu:="" data-blogger-escaped-hr="" data-blogger-escaped-hsmp="" data-blogger-escaped-humidity="" data-blogger-escaped-hz.="" data-blogger-escaped-iec-60601-1-2:="" data-blogger-escaped-ii="" data-blogger-escaped-in="" data-blogger-escaped-include="" data-blogger-escaped-included="" data-blogger-escaped-includes="" data-blogger-escaped-including="" data-blogger-escaped-incorporated="" data-blogger-escaped-incubation="" data-blogger-escaped-incubator="" data-blogger-escaped-independent="" data-blogger-escaped-indicate="" data-blogger-escaped-indication="" data-blogger-escaped-indicative="" data-blogger-escaped-indicator="" data-blogger-escaped-individual="" data-blogger-escaped-inflating="" data-blogger-escaped-inner="" data-blogger-escaped-inorganic="" data-blogger-escaped-input="" data-blogger-escaped-inspection="" data-blogger-escaped-instrument="" data-blogger-escaped-insulated="" data-blogger-escaped-integrated="" data-blogger-escaped-interface="" data-blogger-escaped-interfacing="" data-blogger-escaped-interlock="" data-blogger-escaped-interlocks="" data-blogger-escaped-internal="" data-blogger-escaped-interrupted="" data-blogger-escaped-into="" data-blogger-escaped-intrinsic="" data-blogger-escaped-iron="" data-blogger-escaped-is="" data-blogger-escaped-iso="" data-blogger-escaped-item="" data-blogger-escaped-ix="" data-blogger-escaped-k="" data-blogger-escaped-key.="" data-blogger-escaped-kinetic="" data-blogger-escaped-kit="" data-blogger-escaped-l="" data-blogger-escaped-laboratory="" data-blogger-escaped-lamp="" data-blogger-escaped-lcd="" data-blogger-escaped-ldh="" data-blogger-escaped-leak="" data-blogger-escaped-least="" data-blogger-escaped-less="" data-blogger-escaped-light="" data-blogger-escaped-linearity="" data-blogger-escaped-lis="" data-blogger-escaped-liters="" data-blogger-escaped-local="" data-blogger-escaped-lock="" data-blogger-escaped-locked="" data-blogger-escaped-locks="" data-blogger-escaped-long="" data-blogger-escaped-low="" data-blogger-escaped-lt.="" data-blogger-escaped-lung="" data-blogger-escaped-lymphocyte="" data-blogger-escaped-m="" data-blogger-escaped-made="" data-blogger-escaped-magnetic="" data-blogger-escaped-maintenance-free="" data-blogger-escaped-maintenance="" data-blogger-escaped-manual="" data-blogger-escaped-manufactured="" data-blogger-escaped-manufacturer="" data-blogger-escaped-marked="" data-blogger-escaped-masks="" data-blogger-escaped-mean="" data-blogger-escaped-means="" data-blogger-escaped-measurement="" data-blogger-escaped-medicine="" data-blogger-escaped-meet="" data-blogger-escaped-memory="" data-blogger-escaped-menu="" data-blogger-escaped-mesh="" data-blogger-escaped-meter="" data-blogger-escaped-mg="" data-blogger-escaped-min.="" data-blogger-escaped-minimum="" data-blogger-escaped-minutes.="" data-blogger-escaped-ml.="" data-blogger-escaped-ml="" data-blogger-escaped-mn.="" data-blogger-escaped-mode="" data-blogger-escaped-model="" data-blogger-escaped-monocyte="" data-blogger-escaped-more="" data-blogger-escaped-motor="" data-blogger-escaped-na="" data-blogger-escaped-network="" data-blogger-escaped-nm="" data-blogger-escaped-normally="" data-blogger-escaped-not="" data-blogger-escaped-number="" data-blogger-escaped-obturator="" data-blogger-escaped-of="" data-blogger-escaped-offer="" data-blogger-escaped-offered="" data-blogger-escaped-open="" data-blogger-escaped-operate="" data-blogger-escaped-operating="" data-blogger-escaped-operation="" data-blogger-escaped-operational="" data-blogger-escaped-optical="" data-blogger-escaped-or="" data-blogger-escaped-orifice="" data-blogger-escaped-others="" data-blogger-escaped-outer="" data-blogger-escaped-outpatient="" data-blogger-escaped-over="" data-blogger-escaped-page="" data-blogger-escaped-panel="" data-blogger-escaped-paper="" data-blogger-escaped-parameters="" data-blogger-escaped-parts="" data-blogger-escaped-patient="" data-blogger-escaped-pc="" data-blogger-escaped-pcs="" data-blogger-escaped-pep="" data-blogger-escaped-per="" data-blogger-escaped-performed:="" data-blogger-escaped-performed="" data-blogger-escaped-period="" data-blogger-escaped-permit="" data-blogger-escaped-phase="" data-blogger-escaped-phosphorus="" data-blogger-escaped-plasma="" data-blogger-escaped-plastic="" data-blogger-escaped-platelet="" data-blogger-escaped-please="" data-blogger-escaped-point="" data-blogger-escaped-polysulfone="" data-blogger-escaped-ports="" data-blogger-escaped-position="" data-blogger-escaped-positive="" data-blogger-escaped-possibility="" data-blogger-escaped-possible="" data-blogger-escaped-power="" data-blogger-escaped-pre-selection="" data-blogger-escaped-precise="" data-blogger-escaped-precision-25="" data-blogger-escaped-presenting="" data-blogger-escaped-pressure="" data-blogger-escaped-printer="" data-blogger-escaped-process="" data-blogger-escaped-productions="" data-blogger-escaped-program="" data-blogger-escaped-programmable="" data-blogger-escaped-progress="" data-blogger-escaped-protecting="" data-blogger-escaped-protective="" data-blogger-escaped-protein="" data-blogger-escaped-pt:="" data-blogger-escaped-pt="" data-blogger-escaped-purchaser="" data-blogger-escaped-purging="" data-blogger-escaped-quality="" data-blogger-escaped-r="" data-blogger-escaped-range:="" data-blogger-escaped-range="" data-blogger-escaped-ranges="" data-blogger-escaped-rapid="" data-blogger-escaped-rate:="" data-blogger-escaped-ratio="" data-blogger-escaped-reader="" data-blogger-escaped-reagent="" data-blogger-escaped-reagents="" data-blogger-escaped-red="" data-blogger-escaped-refrigerator="" data-blogger-escaped-regulated="" data-blogger-escaped-regulator="" data-blogger-escaped-relative="" data-blogger-escaped-relevant="" data-blogger-escaped-relief="" data-blogger-escaped-reporting="" data-blogger-escaped-required="" data-blogger-escaped-requirements:="" data-blogger-escaped-requirements="" data-blogger-escaped-resistant="" data-blogger-escaped-results="" data-blogger-escaped-resuscitator="" data-blogger-escaped-retake="" data-blogger-escaped-rolls="" data-blogger-escaped-rotors="" data-blogger-escaped-rpm.="" data-blogger-escaped-rpm="" data-blogger-escaped-rs232="" data-blogger-escaped-rs="" data-blogger-escaped-rubber="" data-blogger-escaped-running="" data-blogger-escaped-s="" data-blogger-escaped-safety="" data-blogger-escaped-samples.="" data-blogger-escaped-samples="" data-blogger-escaped-sampling="" data-blogger-escaped-scale="" data-blogger-escaped-seamless="" data-blogger-escaped-seats="" data-blogger-escaped-second="" data-blogger-escaped-sections="" data-blogger-escaped-see-through="" data-blogger-escaped-selective="" data-blogger-escaped-self-contained="" data-blogger-escaped-self-filling="" data-blogger-escaped-self="" data-blogger-escaped-semi-="" data-blogger-escaped-semi-automated="" data-blogger-escaped-semi="" data-blogger-escaped-separate="" data-blogger-escaped-serum="" data-blogger-escaped-setting="" data-blogger-escaped-shall="" data-blogger-escaped-shelves.="" data-blogger-escaped-shelves="" data-blogger-escaped-shielded="" data-blogger-escaped-should="" data-blogger-escaped-silicon="" data-blogger-escaped-silicone="" data-blogger-escaped-single="" data-blogger-escaped-size="" data-blogger-escaped-software="" data-blogger-escaped-spares="" data-blogger-escaped-specifications:="" data-blogger-escaped-specifications="" data-blogger-escaped-speed="" data-blogger-escaped-stability="" data-blogger-escaped-stainless="" data-blogger-escaped-standard="" data-blogger-escaped-standards.="" data-blogger-escaped-standards:="" data-blogger-escaped-standards="" data-blogger-escaped-start-up="" data-blogger-escaped-start="" data-blogger-escaped-starter="" data-blogger-escaped-steam="" data-blogger-escaped-steel.="" data-blogger-escaped-steel="" data-blogger-escaped-step="" data-blogger-escaped-sterilization="" data-blogger-escaped-sterilizer="" data-blogger-escaped-sterilizing="" data-blogger-escaped-still="" data-blogger-escaped-stirrers="" data-blogger-escaped-stirring="" data-blogger-escaped-stirs="" data-blogger-escaped-stop="" data-blogger-escaped-storage.="" data-blogger-escaped-storage="" data-blogger-escaped-store="" data-blogger-escaped-stored="" data-blogger-escaped-straight="" data-blogger-escaped-straps="" data-blogger-escaped-strips="" data-blogger-escaped-strong="" data-blogger-escaped-supplied="" data-blogger-escaped-supply:="" data-blogger-escaped-supply="" data-blogger-escaped-surface.="" data-blogger-escaped-swing="" data-blogger-escaped-switch="" data-blogger-escaped-system="" data-blogger-escaped-table="" data-blogger-escaped-tap="" data-blogger-escaped-technical="" data-blogger-escaped-temperature:="" data-blogger-escaped-temperature="" data-blogger-escaped-test="" data-blogger-escaped-testing="" data-blogger-escaped-tests="" data-blogger-escaped-than="" data-blogger-escaped-the="" data-blogger-escaped-thermometer="" data-blogger-escaped-thermoprinter="" data-blogger-escaped-thermostat.="" data-blogger-escaped-thermostat="" data-blogger-escaped-thin="" data-blogger-escaped-this="" data-blogger-escaped-three="" data-blogger-escaped-throughput="" data-blogger-escaped-time="" data-blogger-escaped-timer="" data-blogger-escaped-timing="" data-blogger-escaped-to="" data-blogger-escaped-tolerance="" data-blogger-escaped-top="" data-blogger-escaped-total="" data-blogger-escaped-toxic="" data-blogger-escaped-transferring="" data-blogger-escaped-transmit="" data-blogger-escaped-transparent="" data-blogger-escaped-trays="" data-blogger-escaped-triglycerides="" data-blogger-escaped-trips="" data-blogger-escaped-tube="" data-blogger-escaped-tubes="" data-blogger-escaped-two="" data-blogger-escaped-type="" data-blogger-escaped-under="" data-blogger-escaped-unidirectional="" data-blogger-escaped-uniformity="" data-blogger-escaped-unit="" data-blogger-escaped-units="" data-blogger-escaped-up="" data-blogger-escaped-upon="" data-blogger-escaped-upright="" data-blogger-escaped-urea="" data-blogger-escaped-uric="" data-blogger-escaped-urinalysis="" data-blogger-escaped-urine="" data-blogger-escaped-us="" data-blogger-escaped-use="" data-blogger-escaped-used="" data-blogger-escaped-user="" data-blogger-escaped-uv="" data-blogger-escaped-uvc="" data-blogger-escaped-v.="" data-blogger-escaped-v="" data-blogger-escaped-vac="" data-blogger-escaped-vacuum="" data-blogger-escaped-values="" data-blogger-escaped-valve="" data-blogger-escaped-variation="" data-blogger-escaped-various="" data-blogger-escaped-vascular="" data-blogger-escaped-vent="" data-blogger-escaped-ventilation.="" data-blogger-escaped-ventilation="" data-blogger-escaped-version="" data-blogger-escaped-vii="" data-blogger-escaped-viii="" data-blogger-escaped-vinyl="" data-blogger-escaped-visual="" data-blogger-escaped-volume="" data-blogger-escaped-volumes="" data-blogger-escaped-wall="" data-blogger-escaped-washable="" data-blogger-escaped-water="" data-blogger-escaped-wbc="" data-blogger-escaped-well="" data-blogger-escaped-when="" data-blogger-escaped-which="" data-blogger-escaped-whole="" data-blogger-escaped-width="" data-blogger-escaped-window.="" data-blogger-escaped-with="" data-blogger-escaped-within="" data-blogger-escaped-wool.="" data-blogger-escaped-work="" data-blogger-escaped-x="" data-blogger-escaped-z3cn18="" data-blogger-escaped-zero="" data-blogger-escaped-zeroing.="">280nm) with Strong bactericide effectiveness with a maximum
strength situated at 265nm (253.7nm Mercury Spectra )
UVC shall be emitted thru vacuum ionizing Tube ( Weak permeation )
Useful power shall be at least 45% of nominal power (30W)
LifeTime shall be at least 7500 Hrs
Radiation shall be witnessed by means of blue Light emission
Encased in Steel Sheet casing , Epoxy-painted. Ceiling or wall fixation
Power supply:220V.AC50 Hz.
Accessories/Spare Parts:
Ionized Tube : 5
46 Infusion fluid holder
HSMP Equipment & Furniture Component HPIU: HH
equipment specifications[1].doc 16/03/2011 page 22 of 27
20 IV Pole , Mobile on Castors
Adjustable height
2 Hooks
47 Patient trolley
5
Manufactured from chrome plated steel tube
Upholstered fixed Top covered with washable plastic material , Flame Retardant
Complete with side Rails and IV Stand
Mounted on 4 casters 100 mm two of which with brakes
Knock down construction.
Length min. –2000 mm.
Height - 700 mm.
48 Incubator neonate
Nursing incubator mobile on castors
Temperature control by air temperature or skin temperature
Temperature control range 20-37 °C
Display of temperature
Alarms
High and low air temperature
High and low skin temperature
Fan failure
Sensor failure
Power failure
Secondary thermostat
Heater power indicator
At least 2 hand ports
At least 2 tubing ports
Provision for oxygen supplementation
Provision for humidity supplementation
Internal noise < 60 dB(A)
Power supply 220 V 50 Hz
49. Surgical light mobile
Mobile, Stand-type.Max. Light intensity (Ec) up to not less than 110
000 lx or more Color rendering index (general) Ra 94-96 Color
rendering index (red) R9> 90 Light field size approx. 13-25 cm. Co lour
temperature 4200º K (approx.) Working range without focusing
intensity approx.70 – 130 cm Power supply 220 V, 50 Hz Light source 3
x halogen approx. 23 V /50 W Durability of halogen bulbs 1300 h or
more Adjustment of light intensity approx.50 – 100% (0-100%)
Sterilizable handle
50. Splint set required
51. Transportation splint required
52. Stretcher required
54. Tracheostomy set required
55. Fiber bronchoscope
Optical System
Field of view: 100120
Depth of field: 350 mm.
Distal end
Bending Section
Range of tip bending: not less than Up 180, Down 130.
Insertion tube
Outer diameter: not more than 5.0mm
Working length: not less than 600mm.
Total lenhgth: not more than 900mm.
Instrument channel
Inner diameter: not less than 2.0mm.
Light Source
Mobile
Lamp
Type:Quartz halogen.
Output: 150W.
Cooling fan prolongs lamp life not less than 500 hours.
Air/Water functions available
Power consumption:not more than 250 W.
Power Requirements - 220 ± 10% V. 50 Hz. standard network
Weight: not more than 6 kg.
HSMP Equipment & Furniture Component HPIU: HH
equipment specifications[1].doc 16/03/2011 page 23 of 27
56. Arthroscope
Telescope: 1.2mm , 0º direction of view. Trocar Tube: 1.2mm Trocar
spike: 1.2mm Accessories: Adapter with Luer lock male and x-mas tree
Telescope holder Light quid, 3.5m with condenser Trocar Tube for
surgical instruments 2mm Trocar spike 2mm pointed Trocar spike 2mm
blunt Palpation Probe 2 mm straight -4pcs. Palpation Probe 2mm Hook
Type -4pcs. Knife 2mm side cutting -4pcs. Knife 2mm front cutting -
4pcs. Rasp 2mm -4pcs. Biopsy Forceps 2mm -4pcs. Grasping Forceps
2mm -4pcs. Instruments Tray Shaver System
57. Duodenoscope
Optical System Field of View: 80º minimum Direction of View: Lateral,
Rear-view 15º minimum Depth of Field: 5-60 mm minimum Distal End
Outer Diameter: max. 12.0mm Instrument Channel Inner Diameter:
minimum 2.8mm Insertion Tube Outer Diameter: max. 11.0mm
Working Length: minimum 1235mm Bending Section Range of Tip
Bending minimum: UP 120º, Down 90º Right 110º, Left 90º Biopsy
Forceps Minimum Visible Distance: 10 mm from Distal End
58.
Laparoscopy system with
instrumentation for
general surgery
Laparoscopy Surgery Set Including:
Forward-Oblique Telescope 300 deg, diameter 10mm, length 31cm,
including
fiber optic light transmission incorporated. Color code: red.
Forward-Oblique Telescope 300 deg, diameter 5mm, length 29 cm,
including
fiber optic light transmission incorporated. Color code: red.
VERESS Pneumoperitoneum Needle with spring loaded stylet LUER-lock,
length 13 cm.
Trocar 11mm with Cannula and multifunctional valve.
Trocar 6mm with Cannula and multifunctional valve.
Reducer 11/5mm.
Reduction Sleeve, reusable, instrument diameter 3mm, trocar cannula O.D.
6 mm.
Reduction Sleeve, reusable, instrument diameter 10mm, trocar cannula O.D.
15 mm.
Reduction Sleeve, reusable, instrument diameter 5mm,trocar cannula O.D.
11 mm.color code:green
Dissecting and Grasping Forceps, rotating, with connector for unipolar
coagulation size 5mm, length 36 cm.
KELLY and Grasping Forceps, rotating, with connector for unipolar
coagulation size 5mm, length 36 cm.
KELLY Forceps insert, long
MANHES grasping forceps with connector for unipolar
coagulation 5mm, length 36cm
METZENBAUM scissors, rotating, with connector for unipolar
coagulation 5mm, length 36cm,blades curved, double action jaws, length of
blades 12 mm.
Insert METZENBAUM scissors
Hook Scissors, rotating, size 5mm, length 36cm,single action jaws
Coagulating and Dissecting Electrode, L-shaped, with connector pin for
unipolar coagulation size 5mm, length 36 cm
Coagulating and Dissecting Electrode, spatula-sheaped, blunt with
connector pin for unipolar coagulation size 5mm working length
Insert Micro-dissecting scissors
BLAKSSLEY Forceps insert
Suction and coagulation cannula, L-shaper
Take-apart Bipolar Coagulating Forceps
Insert Forceps
Applier for ligating clips (medium large) dismanting, rotating
Pilling-Titanium-Clips, medium-large, sterile box with 16 cartridges,
10clipseach
Electronic CO2 Endoflator, CO2 bottle empty, power supply, 240 VAC, 50
Hz,
CO2/ N2O gas filter for use with Electronic Endoflator
High Frequency Surgery Unit, power supply: 200-240 VAC, including: 400
A Mains Cord Adaptor, unipolar HF- output, 4 mm/5 mm
HSMP Equipment & Furniture Component HPIU: HH
equipment specifications[1].doc 16/03/2011 page 24 of 27
Connecting Cord, for connecting the neutral electrodes
Neutral-Electrode made of conductive silicone, with 1 rubber strap for
fastening contact face, surface area approx. 180 sq.cm.
Two-Pedal Footswitch, for control of coagulation.
Monopolar High Frequency Cord with 5 mm plug for HF-unit, length 300
cm
Suction Bottle, 1.5L.
Bacterial filter
Cold Light Fountain: approx. 175 W power supply: 220-240 VAC, 50 Hz
Fiber Optic Light Cable, size: 4.8 mm, length 180 cm
Endoscopic video system, color system PAL power supply: 240 VAC, 50
Hz.
Cover for use with endoscopic video cameras, size 17 x 242 cm, rolled up.
Sterile, in packs of 15.
Color Monitor, color system PAL, NTSC screen 45 cm power supply: 240
VAC, 50 Hz including: mains cord
Mobile cart for video system: 4 antistatic dual wheels, 2
equipped with locking brakes, 2 fixes shelves, 1 drawer unit with lock, 1 set
of non-sliding stands for units, inclusive integrated small cable conduit in
both vertical boom, 1 camera mount. Power Box socket board with 12
plugs, 12 grounding plugs Approximate Dimensions: cart: 700mm x
1280mm x 686mm (wxhxd) Shelf 630mm x 480mm (wxd) caster diameter:
125 mm
59. EEG
Electroencephalograph- analyzer. Based on the personal computer. Routine
EEG studies with computerized analysis and topographical mapping, indepth
analysis of EEG/EP activity, analysis of epileptiform activity, longterm
EEG monitoring , differential diagnosis 24-EEG inputs. Sensitivity:
0.1 μV to 200 μV. Calibration μV: square (1Hz.) or Sine (5Hz) signal of 5,
10 ,20, 50, 100, 200 and more High frequency filter: >70 Hz at highest
setting) Low frequency filter:
Electrical works
Saturday, January 24, 2015
Monday, January 12, 2015
Distribution Transformer up
boasts of modern design and production technique supported by Computer Aided Design and required software. Local and Foreign latest technology and experienced consultants and experts are continuously placing their contributions behind the development of quality production.
Our distribution transformers are ruggedly designed for outdoor and indoor applications. Transformer cores are made of high permeable cold rolled grain oriented (CRGO) electrical steel insulated on both sides mostly imported from Japan, Europe, and USA. Primary consideration is to reduce no load losses, no load current and noise level- high quality core clamping bolts, channels and supports being used.
All of our transformer winding are made of high conductivity electrolytic copper, pin hole free enamels wire. For the HV winding one or more elementary copper strip or super enamel insulated wire are used and for the LV winding one or more copper strip insulated with pure cellulose paper are used.
The tank and cover made of lamination-free, high quality sheet steel, fabricated of a rigid bottom with an oil drainage outlet. After the required components are welded, the first leakage test is performed on the tank.
Energypac is probably the only company in private sector have facility for sand blasting to protect the derusting in the high humid country like Bangladesh. An active base coat and two finishing coats are applied in every tank and special tank and paints can be incorporated subject to customer requirement.
Energypac uses mostly off load tap changer and on load tap changer as to customer requirement. The tap changers are sourced from the best and proven sources of Europe or best manufacturer. Our tap changers are compact and mounted on the tap of the tank.
Buchholz Relay detects the internal faults and the insulating liquid loss, the alarm contact in the double float design signals oil leakage and/or gases, while a trip contact is activated in the event of severe malfunctions. We provide Buchholz relay as to customer’s requirement.
Thermometer shows the actual oil temperature, alarm and trip signals are provided as protection against over loading and as per requirement of transformer.
After the H. V. and L. V. coils are mounted and centered, the upper yoke is staggered and the core clamping is installed, the cover is also bolted to the upper clamping. The HV winding leads are insulated with paper tube and soldered to the bushing or the tap changer connection. The HV leads are insulated against the upper steel clamping by one press board panel, while another panel fixed the position of the connection wires, the standard wires of the LV bushing are fitted with crimped cable lugs and securely screwed on to the LV winding terminals.
Energypac is the only private company probably in Bangladesh using vacuum drying to extract accumulated humidity from the core coil assembly. Immediately after drying, the transformers are tanked and insulating oil are filled with a vacuum oil filtering machine.
Source : Energy pac
Cast Resin Transformer
Subscribe to:
Posts (Atom)